Product Details
Classification: Physiological Functions of Diagnosis and Monitoring Equipment
Type: Patient Monitor
Certification: CE, EEC, MSDS, ISO13485
Diamond Member Since 2019

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Number of Employees
14
Year of Establishment
2014-02-25
Management System Certification
ISO 9001, ISO 9000, ISO 13485, WRAP
Find Similar Products

Basic Info.

Model NO.
MS-3900
Group
Middle-aged and Old
G.W
2.5 Kg
N.W
1.5 Kg
Transport Package
1 PCS in Carton
Specification
390*310*640mm
Trademark
customized
Origin
China
HS Code
9018193010
Production Capacity
10000PCS/Year

Product Description

(MS-3900) Multiple Display Modes Patient Monitor SpO2, NIBP, Temp Vital Signs Monitor
(MS-3900) Multiple Display Modes Patient Monitor SpO2, NIBP, Temp Vital Signs Monitor(MS-3900) Multiple Display Modes Patient Monitor SpO2, NIBP, Temp Vital Signs Monitor
MS-3900 Vital Signs Monitor
• Built-in memory, store up to 100 hours of patient SpO2 and 500 groups of patient NIBP
•  Perfusion Index (PI) measurement as blood perfusion indicator
•  High brightness cold cathode fluorescent backlight LCD +LED display
•  Advanced analog processing technology prov accurate measurement during low perfusion a movement situation
•  Adjustable audible and visual alarms
•  Built-in large capacity rechargeable lithium ba
•  5V DC power supply for ambulance use

Model Configuration:
Config. S:SpO2, Pulse Rate, Perfusion Index    Config. SN:SpO2, Pulse Rate, NIBP, Perfusion Index  
Optional Config. :  Nellcor SpO2 , Masimo SpO2    
Optional Accessories:  Wall-mounting, Trolley, Neonatal /Pediatric Accessories
Performance Specifications
Dimension and Weight
-210mm*170mm*75mm
- Weight: 1kg(excluding accessories)
Power Supply
-  Voltage: AC100~240V, 50/60HZ
-  Adapter: Output 5VDC, 4A
Display
-  LED + LCD  
-  LCD Size: 85mm * 30mm  
Battery    
-  Type: Rechargeable lithium battery 3.7V/2200mAh
-  Charge Cycle: ≥500 times
- Working time: 3 hours  
Alarm
-  Level: Low, medium and high
-  Indication: Auditory and visual
-  Alarm volume adjustable
System Output  
-  Ethernet Network: standard RJ45 socket * 1
-  USB Port * 1
Trend & Reviewing
-  Trend: 100 hours of patient SpO2  
-  NIBP measurement reviewing: 500 groups
Safety
-  IEC60601-1 Approved, CE marking according to MDD93/42/EEC
- With reference to RoHS Directive 2011/65/EU recasting
Operating Environment
-  Temperature:  5 ~ 40 ºC
-  Humidity:  15% ~ 80% (non-condensing)
-  Atmosphere pressure:  86 KPa ~ 110 Kpa

Transportation and Storage Environment
-  Temperature:   -20~50 ºC
-  Humidity: 10%~90%  (non-condensing)
-  Atmosphere pressure:  86 KPa ~ 110 KPa
SpO2  
-  Measurement Range: 0 ~ 100 %
-  Resolution: 1 %
-  Support Pitch tone and multi-level volume  
-  Accuracy:    80% ~ 100%, ±2 % (adult)
               70% ~ 80%, ±3% (adult)
              0% ~ 69%, unspecified
-  PI range: 0.075%-20%.  
Pulse Rate
-  Measuring and Alarm Range: 25~250bpm
-  Accuracy: ±1% or ±1 bpm, whichever is greater
-  Resolution: 1bpm
Nellcor SpO2 (Option)
-  Measurement Range: 0 ~ 100 %
-  Resolution: 1 %
-  Accuracy:   70% ~ 100%, ±2 % (adult)
             70% ~ 100%, ±3 % (Neonate)
            70% ~ 100%, ±2 % (Low Perfusion)
            0% ~ 69%, unspecified
Pulse Rate
-  Measurement range: 20 ~ 300 bpm
-  Resolution: 1bpm
-  Accuracy: ±3 bpm (20 ~ 250 bpm)
Unspecified (251~300 bpm)

Masimo SpO2 (Option)
-  Measurement Range: 0 ~ 100 %
-  Resolution: 1 %
-  PI range: 0.025-20%
-  Accuracy:   
70% ~ 100%, ±2 % (adult/ pediatric, non-motion conditions)
70% ~ 100%, ±3 % (neonate, non-motion conditions)
70% ~ 100%, ±3 % (motion conditions)
0% ~ 69% unspecified
Pulse Rate
-  Measurement range: 25 ~ 240 bpm
-  Resolution: 1bpm
-  Accuracy: ±3 bpm (non-motion condition)
NIBP
-  Method: Oscillometric
-  Measure mode: Manual, Auto, STAT
-  Measure Interval in AUTO Mode 1~480 min
-  STAT mode cycle time: Keep 5 minutes, at 5 seconds interval
-  Measure and Alarm Range:  
Adult:      SYS:  30 ~ 270 mmHg
           DIA:  10 ~ 220 mmHg      
           MEAN:  20 ~ 235 mmHg  
Pediatric:    SYS:  40 ~ 235 mmHg
DIA:  10 ~ 220 mmHg      
MEAN:  20 ~ 225 mmHg
Neonate:     SYS:  30 ~ 135 mmHg
DIA:  10 ~ 110 mmHg      
MEAN:  20 ~ 125 mmHg
-  Static pressure accuracy: ±3mmHg
-  Resolution: 1mmHg
-  Accuracy: Maximum Mean error ±5mmHg            
          Maximum Standard deviation ≤8mmHg
- Over  pressure Protection: Dual protection via software & hardware



 