Appliance: | Biochemistry |
---|---|
Ray Type: | X-Ray Spectrometer |
Gaspath System: | Single |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
MS-T300 Fully Automatic Biochemistry Analyzer
Characters:
Equipment type: Fully automatic discrete, STAT priority
Analysis rate Constant speed: 300T/H(without ISE)
Test principle: Colorimetry, turbidimetry
Analysis method: End-point, 2 kinetics, fixed-time, etc. support single/double wavelength, linear and non-linear calibration
Sample System
Sample Tray: 60 sample positions, random access of all samples
Sample Volume: 2-50ul, 0.1ul increment
Sample probe: Liquid level detection, the remaining detection, collision protection etc.
Sample Containers: Sample cups, Micro tube, Original blood-collection tube, etc.
Reagent System
Reagent Tray: 45 reagent positions
Reagent Volume: R1:25-300ul, R2:10-150ul, 0.1ul increment
Reagent Probe Liquid level detection, the remaining detection, collision protection etc.
Reagent Refrigeration: 4-15ºC, independent power supply, 24-hour non-stop refrigeration
Reaction System
Reaction Tray: Rotating tray, 120 cuvettes
Cuvette: Reusable high quality cuvette, Optical diameter 6mm
Reaction Volume: 180-500ul
Incubation syste: Bi-directional liquid-air device
Temperature: 37ºC±0.1ºC with real-time display
Optical System
Light source :12V35W halogen lamps, over 2000 hours
Photoelectron road: after spectrophotometry
Wavelength: 340,405,450,510,546,578,630,700nm
Absorbance range: 0-3.500Abs
Resolution: 0.0001Abs
Operating System
PC operating system: Windows XP or above
Analysis control software: English version graphical operating software
Report printing: Report formats support the user-defined mode, QC and state information etc.
System connection: RS-232
Working Conditions
Power supply: AC 220V±22V, 50Hz±1Hz, power 1.5KVA
Ambient Temperature: 15-30 ºC
Humidity: 35-80%
Water consumption: 7L/Hour
Dimension Bench top: 750mm*520mm*600mm