Type: | Anaesthesia Machine |
---|---|
Kind: | Surgery |
Certification: | CE, FDA, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
MS-600F Anesthesia Machine
Special Feature
*5.5"SNT screen,two waveforms are showed at the same time
*Metal frame,with plastic spray technique,durable and lasting
*Single vaporizer station,selectatec compatible
*Classical flowmeter of 4 tubes
*Could be easily upgrated with functions:eg adding Multi-parameter Monitor,CO2 Monitor,AMIS System
*Working Mode:Semi~open,Semi~closed or Closed system
*Gas Supply Pressure:O2:0.27Mpa~0.55Mpa;N2O:0.27Mpa~0.55Mpa
*Flow Meter
4 Tube:2xO2 Tube,2xN2O Tube,
O2 Tube:0L/min~1L/min~10L/min
N2O Tube: 0L/min~1L/min~10L/min
O2 and N2O linkage Device,Keep Oxygen concentratorion not lower than 25%
*Oxygen Flush
25L/min~75L/min
Safety Function
*When airway pressure reach 6kPa,APL valve will automatically release gas
*When airway pressure reach 6kPa,the bellow will stop sending gas to patient
*When airway pressure reach 6kPa,the bellows
*Monitoring PARAMETERS(UNDER Manual and Ventilation)
·Tidal Volume
·Minute Ventilation
·Respiration Mode
·FiO2 (Optional)
·Waveforms:P-t,V-t
*Alarm Parameters(Audible and/or visual signal)
·Airway pressure high/low limit
·FiO2high/low limit(Optional)
·Low O2 Supply Pressure(When O2 Supply Pressure is lower than 0.15Mpa,will arouse Audible Alarm)
·N2O automatic cut off alarm
Anesthesia Ventilator
*Patient type:Adult,Pedtric
*Driven Mode:Pneumatically Driven and Electronically Controlled
*Respirateion Mode:IPPV,SIMV,Sign,Manual
*Tidal Volume:50~1600ml
*Respiratory Frequency:1~65bpm
*I:E Ratio:8:1~1:8
*Pressure Setting:1~60cmH2O
*Trigger Pressure:-20~20cmH2O
Circle Absorber
*Upgraded integrated circle absorber,with twins large volume CO2 canisters,high CO2 absoring effiiciency
·With no-return respiratory valve and expiratory valve on it avoid gas back flowing
·With APL valve on it,the airway pressure could be adjusted freely
·With "Manual/Ventilation"switch on it,easily and quickly change
Electrical Specification
*Mains:100~240V,50/60Hz
*Battery:12V DC,minimum 120min(if with)
*Power consumption≤50W
Optional
*Yoke Device*Patient Monitor*CO2 Monitor*AMIS*AGSS*FiO2