Classification: | Etco2 |
---|---|
Type: | Etco2 Sensor |
Certification: | ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Transducer Type | Side stream CO2 module(MS-S30) and Mainstream CO2 sensor(MS-M60) |
Principle of Operation | Non-dispersive infra-Red (NDIR) |
Initialization Time | Capnogram displayed in less than 6 seconds(side stream) and 2 senconds (mainstream) |
Reaching full specifications within 2 | |
CO2 Measure Range | 0 - 150mmHg 0 - 20.0% 0- 20kPa |
CO2 Measure Resolution | 0.1mmHg(side stream ), 0.01mmHg(mainstream ) |
Co2 Accuracy | 0 - 40 mmHg ±2 mmHg 41 - 70 mmHg ±5% of reading 71 - 100 mmHg ±8% of reading 101-150 mmHg ±10% of reading |
Respiratory Rate |
Range 2 - 150 RPM |
Accuracy ±1 breath |
|
Calibration |
No routine user calibration required. An airway adapter zero is required when insert an airway adapter |
Compensation |
Barometer Pressure 450 - 850 mmHg. Operator selectable O2 , N2O, He compensation. |
Power Consumption |
5.0 VDC ±5%, power consumption less than 1.3W typical |
Temperature / Humidity | Operating: 0 to 50 ºC, 10 to 90% RH |
Storage: -40 to 70ºC, <90% RH |
|
Data Interface | RS232/TTL, 19200 baud rate, standard N-8-1. Fully compatible with Respironics/Masimo CO2 module's communication protocol |