Ms-M60 Mainstream Etco2 Sensor

Product Details
Classification: Etco2
Type: Etco2 Sensor
Certification: ISO13485
Diamond Member Since 2019

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Number of Employees
14
Year of Establishment
2014-02-25
Management System Certification
ISO 9001, ISO 9000, ISO 13485, WRAP
  • Ms-M60 Mainstream Etco2 Sensor
  • Ms-M60 Mainstream Etco2 Sensor
  • Ms-M60 Mainstream Etco2 Sensor
  • Ms-M60 Mainstream Etco2 Sensor
  • Ms-M60 Mainstream Etco2 Sensor
  • Ms-M60 Mainstream Etco2 Sensor
Find Similar Products

Basic Info.

Model NO.
MS-M60
Group
Adult
Transport Package
1PC/Small Carton
Trademark
customized
Origin
China
HS Code
9018193010
Production Capacity
100000

Product Description

Ms-M60 Mainstream Etco2 SensorMs-M60 Mainstream Etco2 SensorMs-M60 Mainstream Etco2 SensorMs-M60 Mainstream Etco2 SensorMs-M60 Mainstream Etco2 Sensor
MS-M60 Mainstream ETCO2 Sensor

Product Introduction
ETCO2 module is the ideal capnography solution. It is plug-in and play. The sensor features advanced technology for measuring realtime CO2 concentration,  respiration  rate, end-tidal CO2 concentration, inspired CO2 concentration. It resist the vapour better with patented technology, make the measured result more accurate.
Annlication Field
Monitoring the patient's respiratory status

Assisting in determining when to intubate or extubate

Verification of ET-tube placement

Alarming when accidental extubation occurs

Ventilator  disconnect detection

Verification of ventilation during transportation


Technical Specifications
Transducer Type Side stream CO2 module(MS-S30) and Mainstream CO2 sensor(MS-M60)
Principle of Operation Non-dispersive infra-Red (NDIR)
 
Initialization Time Capnogram displayed in less than 6 seconds(side stream) and 2 senconds (mainstream)
 
Reaching full specifications within 2
CO2 Measure Range 0 - 150mmHg
0 - 20.0%
0- 20kPa
 
CO2 Measure Resolution 0.1mmHg(side stream ), 0.01mmHg(mainstream )
 
Co2 Accuracy 0 - 40 mmHg ±2 mmHg
41 - 70  mmHg ±5% of reading
71 - 100 mmHg ±8% of reading
101-150 mmHg ±10% of reading
 
Respiratory Rate
 

Range   2 - 150 RPM
 
Accuracy  ±1 breath
 
Calibration
 
No routine user calibration required. An airway adapter zero is required when insert an airway adapter
 
Compensation
 
Barometer Pressure 450 - 850 mmHg.
Operator selectable O2 , N2O, He compensation.
 
Power Consumption
 
5.0 VDC ±5%, power consumption less than 1.3W typical

 
Temperature / Humidity Operating: 0 to 50 ºC, 10 to 90% RH
 
Storage: -40 to 70ºC, <90% RH
 
Data Interface RS232/TTL, 19200 baud rate, standard N-8-1.
Fully compatible with Respironics/Masimo CO2 module's communication protocol
 


 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier